· 8 years ago · Apr 27, 2018, 10:36 PM
1E 27-Apr-2018 15:31:36 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
2E 27-Apr-2018 15:22:24 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
3E 27-Apr-2018 15:22:24 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
4E 27-Apr-2018 15:21:16 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
5E 27-Apr-2018 15:21:16 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
6E 27-Apr-2018 15:19:40 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
7E 27-Apr-2018 15:19:02 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
8E 27-Apr-2018 15:18:32 1 CAP_DVF 0 An unspecified internal error has occurred.
9 Time: 27.4.2018, 15:18:32, Line: 271, File: \MedCom\src\3D\Visualization\FrontEnd\Editor\src/Cap3dFEEditorDlgInit.cpp, Process: CapGM (7816),
10 Text: "It exists no valid Editor license."
11E 27-Apr-2018 15:17:54 34 MRI_PER 0 Scanner is booting.
12 Time: 27.4.2018, 15:17:54, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6984)
13E 27-Apr-2018 15:17:51 35 MRI_PER 0 Scanner is not online (might be switched off).
14 Time: 27.4.2018, 15:17:51, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6984)
15E 27-Apr-2018 15:17:51 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
16 Time: 27.4.2018, 15:17:50, Process: CsaGenericMain (6840)
17E 27-Apr-2018 15:17:43 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection.
18 Time: 27.4.2018, 15:17:43, Process: CsaGenericMain (6568)
19E 27-Apr-2018 15:17:41 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
20 Time: 27.4.2018, 15:17:41, Process: CsaGenericMain (6568)
21E 27-Apr-2018 15:16:40 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
22E 27-Apr-2018 15:16:31 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
23E 27-Apr-2018 15:16:31 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
24E 27-Apr-2018 15:16:30 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
25E 27-Apr-2018 15:16:30 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
26E 27-Apr-2018 15:16:28 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
27E 27-Apr-2018 15:16:28 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
28E 27-Apr-2018 15:16:27 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
29E 27-Apr-2018 15:16:27 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
30E 27-Apr-2018 15:16:26 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
31E 27-Apr-2018 15:16:26 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
32E 27-Apr-2018 15:16:20 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
33E 27-Apr-2018 15:16:18 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
34E 27-Apr-2018 15:16:18 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
35E 27-Apr-2018 15:16:17 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
36E 27-Apr-2018 15:16:17 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
37E 27-Apr-2018 15:16:16 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
38E 27-Apr-2018 15:16:10 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
39E 27-Apr-2018 15:16:10 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
40E 27-Apr-2018 15:16:09 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
41E 27-Apr-2018 15:15:45 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Services\TmFilter\Parameters' with value 'Cue' by process C:\Program Files (x86)\Trend Micro\OfficeScan Client\NTRTScan.exe (Process Id: 2160, User: NT AUTHORITY\SYSTEM).
42E 27-Apr-2018 15:15:45 20 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify file 'c:\program files\mcafee\solidcore\sadmin.exe' by process C:\WINNT\system32\csrss.exe (Process Id: 464, User: NT AUTHORITY\SYSTEM).
43E 27-Apr-2018 15:15:44 20 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify file 'c:\program files\mcafee\solidcore\sadmin.exe' by process C:\WINNT\system32\csrss.exe (Process Id: 464, User: NT AUTHORITY\SYSTEM).
44E 27-Apr-2018 15:15:36 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'ctfmon.exe' by process C:\WINNT\SysWOW64\ctfmon.exe (Process Id: 1900, User: MRC31210\meduser).
45E 27-Apr-2018 15:15:35 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'internat.exe' by process C:\WINNT\system32\ctfmon.exe (Process Id: 1756, User: MRC31210\meduser).
46E 27-Apr-2018 15:15:35 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'ctfmon.exe' by process C:\WINNT\system32\ctfmon.exe (Process Id: 1756, User: MRC31210\meduser).
47E 27-Apr-2018 15:15:28 50 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_LOCAL_MACHINE\SOFTWARE\Classes\Wow6432Node' by process C:\WINNT\system32\csrss.exe (Process Id: 464, User: NT AUTHORITY\SYSTEM).
48E 27-Apr-2018 15:11:58 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
49E 27-Apr-2018 15:11:57 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
50E 27-Apr-2018 15:11:57 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
51E 27-Apr-2018 15:11:56 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
52E 27-Apr-2018 15:11:56 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
53E 27-Apr-2018 15:11:31 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
54E 27-Apr-2018 15:11:24 32 MRI_COS 0 Connection to Image Reconstruction System has been lost.
55 Time: 27.4.2018, 15:06:38, Line: 876, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4852)
56E 27-Apr-2018 15:11:23 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
57E 27-Apr-2018 15:11:13 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details.
58 Time: 27.4.2018, 15:11:13, Line: 156, File: \n4\pkg\MrApplications\MrExamUI\MrOnlineDisplay\MrDisplayerSvc\MrDisplayerDBSession.cpp, Process: CsaGenericMain (4212),
59 Text: Failed to end session
60 [LOG] An error occurred during accessing db. See logfile for details.
61E 27-Apr-2018 15:11:13 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
62E 27-Apr-2018 15:10:51 35 MRI_PER 29 Scanner is not online (might be switched off).
63 Time: 27.4.2018, 15:10:17, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
64E 27-Apr-2018 15:10:51 1 MRI_PER 28 Scanner is not online. Wait until booting has finished, or switch on Scanner.
65 Time: 27.4.2018, 15:10:17, Process: CsaGenericMain (6900)
66E 27-Apr-2018 15:10:51 15 MRI_PER 27 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
67 Time: 27.4.2018, 15:10:17, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
68E 27-Apr-2018 15:10:51 29 MRI_SLO 26 RF Power Amplifier error: RF pulse too long or pulse time interval too short. Retry, check sequence and/or sequence parameter. If the problem still persists, please call Siemens Service.
69 Time: 27.4.2018, 15:10:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
70E 27-Apr-2018 15:10:51 28 MRI_SLO 25 RF Power Amplifier error: Temperature of the RFPA too high. Retry after 5 minutes. If the problem still persists, please call Siemens Service.
71 Time: 27.4.2018, 15:10:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
72E 27-Apr-2018 15:10:51 27 MRI_SLO 24 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
73 Time: 27.4.2018, 15:10:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
74E 27-Apr-2018 15:10:51 207 MRI_SUP 23 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
75 Time: 27.4.2018, 15:10:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
76E 27-Apr-2018 15:10:51 27 MRI_PAD 22 RF Power Absorption Limiter error: RF limit for 5th averaging stage exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
77 Time: 27.4.2018, 15:08:48, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
78E 27-Apr-2018 15:10:51 71 MRI_GXX 21 Gradient Power Amplifier error: Regulation error on Z-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
79 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
80E 27-Apr-2018 15:10:51 21 MRI_GXX 20 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
81 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
82E 27-Apr-2018 15:10:51 2004 MRI_GXX 19 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
83 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
84E 27-Apr-2018 15:10:51 2000 MRI_GXX 18 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
85 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
86E 27-Apr-2018 15:10:51 71 MRI_GXX 17 Gradient Power Amplifier error: Regulation error on Z-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
87 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
88E 27-Apr-2018 15:10:51 21 MRI_GXX 16 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
89 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
90E 27-Apr-2018 15:10:51 2004 MRI_GXX 15 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
91 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
92E 27-Apr-2018 15:10:51 9 MRI_STD 14 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
93 Time: 27.4.2018, 15:08:45, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
94E 27-Apr-2018 15:10:51 55 MRI_GXX 13 Gradient Power Amplifier error: Regulation error on Y-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
95 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
96E 27-Apr-2018 15:10:51 21 MRI_GXX 12 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
97 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
98E 27-Apr-2018 15:10:51 2004 MRI_GXX 11 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
99 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
100E 27-Apr-2018 15:10:51 2000 MRI_GXX 10 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
101 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
102E 27-Apr-2018 15:10:51 55 MRI_GXX 9 Gradient Power Amplifier error: Regulation error on Y-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
103 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
104E 27-Apr-2018 15:10:51 21 MRI_GXX 8 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
105 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
106E 27-Apr-2018 15:10:51 2004 MRI_GXX 7 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
107 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
108E 27-Apr-2018 15:10:51 9 MRI_STD 6 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
109 Time: 27.4.2018, 15:08:41, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
110E 27-Apr-2018 15:10:51 39 MRI_GXX 5 Gradient Power Amplifier error: Regulation error on X-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
111 Time: 27.4.2018, 15:08:40, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
112E 27-Apr-2018 15:10:51 21 MRI_GXX 4 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
113 Time: 27.4.2018, 15:08:40, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
114E 27-Apr-2018 15:10:51 2004 MRI_GXX 3 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
115 Time: 27.4.2018, 15:08:40, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
116E 27-Apr-2018 15:10:51 2000 MRI_GXX 2 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
117 Time: 27.4.2018, 15:08:40, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
118E 27-Apr-2018 15:10:51 39 MRI_GXX 1 Gradient Power Amplifier error: Regulation error on X-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
119 Time: 27.4.2018, 15:08:39, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6900)
120E 27-Apr-2018 15:10:41 35 MRI_PER 30 Scanner is not online (might be switched off).
121 Time: 27.4.2018, 15:10:17, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
122E 27-Apr-2018 15:10:41 15 MRI_PER 29 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
123 Time: 27.4.2018, 15:10:17, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
124E 27-Apr-2018 15:10:41 29 MRI_SLO 28 RF Power Amplifier error: RF pulse too long or pulse time interval too short. Retry, check sequence and/or sequence parameter. If the problem still persists, please call Siemens Service.
125 Time: 27.4.2018, 15:10:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
126E 27-Apr-2018 15:10:41 28 MRI_SLO 27 RF Power Amplifier error: Temperature of the RFPA too high. Retry after 5 minutes. If the problem still persists, please call Siemens Service.
127 Time: 27.4.2018, 15:10:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
128E 27-Apr-2018 15:10:41 27 MRI_SLO 26 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
129 Time: 27.4.2018, 15:10:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
130E 27-Apr-2018 15:10:41 207 MRI_SUP 25 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
131 Time: 27.4.2018, 15:10:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
132E 27-Apr-2018 15:10:41 27 MRI_PAD 24 RF Power Absorption Limiter error: RF limit for 5th averaging stage exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
133 Time: 27.4.2018, 15:08:48, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
134E 27-Apr-2018 15:10:41 71 MRI_GXX 23 Gradient Power Amplifier error: Regulation error on Z-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
135 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
136E 27-Apr-2018 15:10:41 21 MRI_GXX 22 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
137 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
138E 27-Apr-2018 15:10:41 2004 MRI_GXX 21 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
139 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
140E 27-Apr-2018 15:10:41 2000 MRI_GXX 20 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
141 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
142E 27-Apr-2018 15:10:41 71 MRI_GXX 19 Gradient Power Amplifier error: Regulation error on Z-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
143 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
144E 27-Apr-2018 15:10:41 21 MRI_GXX 18 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
145 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
146E 27-Apr-2018 15:10:41 2004 MRI_GXX 17 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
147 Time: 27.4.2018, 15:08:47, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
148E 27-Apr-2018 15:10:41 9 MRI_STD 16 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
149 Time: 27.4.2018, 15:08:45, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
150E 27-Apr-2018 15:10:41 55 MRI_GXX 15 Gradient Power Amplifier error: Regulation error on Y-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
151 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
152E 27-Apr-2018 15:10:41 21 MRI_GXX 14 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
153 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
154E 27-Apr-2018 15:10:41 2004 MRI_GXX 13 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
155 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
156E 27-Apr-2018 15:10:41 2000 MRI_GXX 12 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
157 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
158E 27-Apr-2018 15:10:41 55 MRI_GXX 11 Gradient Power Amplifier error: Regulation error on Y-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
159 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
160E 27-Apr-2018 15:10:41 21 MRI_GXX 10 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
161 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
162E 27-Apr-2018 15:10:41 2004 MRI_GXX 9 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
163 Time: 27.4.2018, 15:08:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
164E 27-Apr-2018 15:10:41 9 MRI_STD 8 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
165 Time: 27.4.2018, 15:08:41, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
166E 27-Apr-2018 15:10:41 39 MRI_GXX 7 Gradient Power Amplifier error: Regulation error on X-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
167 Time: 27.4.2018, 15:08:40, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
168E 27-Apr-2018 15:10:41 21 MRI_GXX 6 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
169 Time: 27.4.2018, 15:08:40, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
170E 27-Apr-2018 15:10:41 2004 MRI_GXX 5 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
171 Time: 27.4.2018, 15:08:40, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
172E 27-Apr-2018 15:10:41 2000 MRI_GXX 4 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
173 Time: 27.4.2018, 15:08:40, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
174E 27-Apr-2018 15:10:41 39 MRI_GXX 3 Gradient Power Amplifier error: Regulation error on X-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
175 Time: 27.4.2018, 15:08:39, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
176E 27-Apr-2018 15:10:41 21 MRI_GXX 2 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
177 Time: 27.4.2018, 15:08:39, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
178E 27-Apr-2018 15:10:41 2004 MRI_GXX 1 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
179 Time: 27.4.2018, 15:08:39, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
180E 27-Apr-2018 15:10:17 35 MRI_PER 0 Scanner is not online (might be switched off).
181 Time: 27.4.2018, 15:10:17, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
182E 27-Apr-2018 15:10:17 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
183 Time: 27.4.2018, 15:10:17, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
184E 27-Apr-2018 15:10:06 207 MRI_SUP 0 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
185 Time: 27.4.2018, 15:10:06, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
186E 27-Apr-2018 15:09:31 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
187E 27-Apr-2018 15:08:53 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
188E 27-Apr-2018 15:08:23 1 CAP_DVF 0 An unspecified internal error has occurred.
189 Time: 27.4.2018, 15:08:23, Line: 271, File: \MedCom\src\3D\Visualization\FrontEnd\Editor\src/Cap3dFEEditorDlgInit.cpp, Process: CapGM (7856),
190 Text: "It exists no valid Editor license."
191E 27-Apr-2018 15:07:45 34 MRI_PER 0 Scanner is booting.
192 Time: 27.4.2018, 15:07:45, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
193E 27-Apr-2018 15:07:42 35 MRI_PER 0 Scanner is not online (might be switched off).
194 Time: 27.4.2018, 15:07:42, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6952)
195E 27-Apr-2018 15:07:42 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
196 Time: 27.4.2018, 15:07:41, Process: CsaGenericMain (6812)
197E 27-Apr-2018 15:07:34 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection.
198 Time: 27.4.2018, 15:07:34, Process: CsaGenericMain (6540)
199E 27-Apr-2018 15:07:33 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
200 Time: 27.4.2018, 15:07:33, Process: CsaGenericMain (6540)
201E 27-Apr-2018 15:06:34 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
202E 27-Apr-2018 15:06:27 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
203E 27-Apr-2018 15:06:27 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
204E 27-Apr-2018 15:06:27 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
205E 27-Apr-2018 15:06:26 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
206E 27-Apr-2018 15:06:24 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
207E 27-Apr-2018 15:06:24 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
208E 27-Apr-2018 15:06:23 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
209E 27-Apr-2018 15:06:23 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
210E 27-Apr-2018 15:06:22 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
211E 27-Apr-2018 15:06:22 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
212E 27-Apr-2018 15:06:14 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
213E 27-Apr-2018 15:06:14 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
214E 27-Apr-2018 15:06:14 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
215E 27-Apr-2018 15:06:13 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
216E 27-Apr-2018 15:06:12 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
217E 27-Apr-2018 15:06:03 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
218E 27-Apr-2018 15:05:59 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
219E 27-Apr-2018 15:05:34 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Services\TmFilter\Parameters' with value 'Cue' by process C:\Program Files (x86)\Trend Micro\OfficeScan Client\NTRTScan.exe (Process Id: 2224, User: NT AUTHORITY\SYSTEM).
220E 27-Apr-2018 15:05:33 20 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify file 'c:\program files\mcafee\solidcore\sadmin.exe' by process C:\WINNT\system32\csrss.exe (Process Id: 464, User: NT AUTHORITY\SYSTEM).
221E 27-Apr-2018 15:05:32 20 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify file 'c:\program files\mcafee\solidcore\sadmin.exe' by process C:\WINNT\system32\csrss.exe (Process Id: 464, User: NT AUTHORITY\SYSTEM).
222E 27-Apr-2018 15:05:22 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'ctfmon.exe' by process C:\WINNT\SysWOW64\ctfmon.exe (Process Id: 1888, User: MRC31210\meduser).
223E 27-Apr-2018 15:05:22 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'internat.exe' by process C:\WINNT\system32\ctfmon.exe (Process Id: 1760, User: MRC31210\meduser).
224E 27-Apr-2018 15:05:21 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'ctfmon.exe' by process C:\WINNT\system32\ctfmon.exe (Process Id: 1760, User: MRC31210\meduser).
225E 27-Apr-2018 15:05:13 50 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_LOCAL_MACHINE\SOFTWARE\Classes\Wow6432Node' by process C:\WINNT\system32\csrss.exe (Process Id: 464, User: NT AUTHORITY\SYSTEM).
226E 27-Apr-2018 15:02:23 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
227E 27-Apr-2018 15:02:22 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
228E 27-Apr-2018 15:02:22 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
229E 27-Apr-2018 15:02:21 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
230E 27-Apr-2018 15:01:42 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
231E 27-Apr-2018 15:01:28 32 MRI_COS 3 Connection to Image Reconstruction System has been lost.
232 Time: 27.4.2018, 14:48:29, Line: 876, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4996)
233E 27-Apr-2018 15:01:28 11 MRI_COS 1 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords.
234 Time: 27.4.2018, 07:17:20, Line: 793, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4996)
235E 27-Apr-2018 15:01:28 32 MRI_COS 0 Connection to Image Reconstruction System has been lost.
236 Time: 27.4.2018, 07:17:20, Line: 862, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4996)
237E 27-Apr-2018 15:01:18 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details.
238 Time: 27.4.2018, 15:01:18, Line: 156, File: \n4\pkg\MrApplications\MrExamUI\MrOnlineDisplay\MrDisplayerSvc\MrDisplayerDBSession.cpp, Process: CsaGenericMain (4328),
239 Text: Failed to end session
240 [LOG] An error occurred during accessing db. See logfile for details.
241E 27-Apr-2018 15:01:18 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
242E 27-Apr-2018 15:01:16 1 MRI_PER 1 Scanner is not online. Wait until booting has finished, or switch on Scanner.
243 Time: 27.4.2018, 15:00:32, Process: CsaGenericMain (5292)
244E 27-Apr-2018 15:01:16 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
245 Time: 27.4.2018, 14:55:22, Process: CsaGenericMain (5292)
246E 27-Apr-2018 15:00:37 1 MRI_PER 30 Scanner is not online. Wait until booting has finished, or switch on Scanner.
247 Time: 27.4.2018, 15:00:32, Process: CsaGenericMain (6592)
248E 27-Apr-2018 15:00:37 32 MRI_COH 28 Patient table error: Table position check failed Please retry the measurement. Please save System Log-Files immediately by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service.
249 Time: 27.4.2018, 14:59:15, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
250E 27-Apr-2018 15:00:37 27 MRI_PAD 27 RF Power Absorption Limiter error: RF limit for 5th averaging stage exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
251 Time: 27.4.2018, 14:57:34, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
252E 27-Apr-2018 15:00:37 71 MRI_GXX 26 Gradient Power Amplifier error: Regulation error on Z-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
253 Time: 27.4.2018, 14:57:34, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
254E 27-Apr-2018 15:00:37 21 MRI_GXX 25 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
255 Time: 27.4.2018, 14:57:34, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
256E 27-Apr-2018 15:00:37 2004 MRI_GXX 24 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
257 Time: 27.4.2018, 14:57:34, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
258E 27-Apr-2018 15:00:37 2000 MRI_GXX 23 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
259 Time: 27.4.2018, 14:57:34, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
260E 27-Apr-2018 15:00:37 71 MRI_GXX 22 Gradient Power Amplifier error: Regulation error on Z-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
261 Time: 27.4.2018, 14:57:33, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
262E 27-Apr-2018 15:00:37 21 MRI_GXX 21 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
263 Time: 27.4.2018, 14:57:33, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
264E 27-Apr-2018 15:00:37 2004 MRI_GXX 20 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
265 Time: 27.4.2018, 14:57:33, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
266E 27-Apr-2018 15:00:37 9 MRI_STD 19 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
267 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
268E 27-Apr-2018 15:00:37 55 MRI_GXX 18 Gradient Power Amplifier error: Regulation error on Y-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
269 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
270E 27-Apr-2018 15:00:37 21 MRI_GXX 17 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
271 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
272E 27-Apr-2018 15:00:37 2004 MRI_GXX 16 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
273 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
274E 27-Apr-2018 15:00:37 2000 MRI_GXX 15 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
275 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
276E 27-Apr-2018 15:00:37 55 MRI_GXX 14 Gradient Power Amplifier error: Regulation error on Y-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
277 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
278E 27-Apr-2018 15:00:37 21 MRI_GXX 13 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
279 Time: 27.4.2018, 14:57:29, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
280E 27-Apr-2018 15:00:37 2004 MRI_GXX 12 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
281 Time: 27.4.2018, 14:57:29, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
282E 27-Apr-2018 15:00:37 9 MRI_STD 11 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
283 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
284E 27-Apr-2018 15:00:37 39 MRI_GXX 10 Gradient Power Amplifier error: Regulation error on X-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
285 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
286E 27-Apr-2018 15:00:37 21 MRI_GXX 9 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
287 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
288E 27-Apr-2018 15:00:37 2004 MRI_GXX 8 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
289 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
290E 27-Apr-2018 15:00:37 2000 MRI_GXX 7 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
291 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
292E 27-Apr-2018 15:00:37 39 MRI_GXX 6 Gradient Power Amplifier error: Regulation error on X-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
293 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
294E 27-Apr-2018 15:00:37 21 MRI_GXX 5 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
295 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
296E 27-Apr-2018 15:00:37 2004 MRI_GXX 4 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
297 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
298E 27-Apr-2018 15:00:37 9 MRI_STD 3 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
299 Time: 27.4.2018, 14:57:22, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
300E 27-Apr-2018 15:00:37 801 MRI_COH 2 Scanner control error: Measurement is blocked by the MSUP supervisory test. Retry after the supervisory test is finished.
301 Time: 27.4.2018, 14:57:21, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6592)
302E 27-Apr-2018 15:00:32 32 MRI_COH 30 Patient table error: Table position check failed Please retry the measurement. Please save System Log-Files immediately by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service.
303 Time: 27.4.2018, 14:59:15, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
304E 27-Apr-2018 15:00:32 27 MRI_PAD 29 RF Power Absorption Limiter error: RF limit for 5th averaging stage exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
305 Time: 27.4.2018, 14:57:34, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
306E 27-Apr-2018 15:00:32 71 MRI_GXX 28 Gradient Power Amplifier error: Regulation error on Z-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
307 Time: 27.4.2018, 14:57:34, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
308E 27-Apr-2018 15:00:32 21 MRI_GXX 27 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
309 Time: 27.4.2018, 14:57:34, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
310E 27-Apr-2018 15:00:32 2004 MRI_GXX 26 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
311 Time: 27.4.2018, 14:57:34, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
312E 27-Apr-2018 15:00:32 2000 MRI_GXX 25 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
313 Time: 27.4.2018, 14:57:34, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
314E 27-Apr-2018 15:00:32 71 MRI_GXX 24 Gradient Power Amplifier error: Regulation error on Z-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
315 Time: 27.4.2018, 14:57:33, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
316E 27-Apr-2018 15:00:32 21 MRI_GXX 23 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
317 Time: 27.4.2018, 14:57:33, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
318E 27-Apr-2018 15:00:32 2004 MRI_GXX 22 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
319 Time: 27.4.2018, 14:57:33, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
320E 27-Apr-2018 15:00:32 9 MRI_STD 21 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
321 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
322E 27-Apr-2018 15:00:32 55 MRI_GXX 20 Gradient Power Amplifier error: Regulation error on Y-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
323 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
324E 27-Apr-2018 15:00:32 21 MRI_GXX 19 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
325 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
326E 27-Apr-2018 15:00:32 2004 MRI_GXX 18 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
327 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
328E 27-Apr-2018 15:00:32 2000 MRI_GXX 17 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
329 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
330E 27-Apr-2018 15:00:32 55 MRI_GXX 16 Gradient Power Amplifier error: Regulation error on Y-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
331 Time: 27.4.2018, 14:57:30, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
332E 27-Apr-2018 15:00:32 21 MRI_GXX 15 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
333 Time: 27.4.2018, 14:57:29, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
334E 27-Apr-2018 15:00:32 2004 MRI_GXX 14 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
335 Time: 27.4.2018, 14:57:29, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
336E 27-Apr-2018 15:00:32 9 MRI_STD 13 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
337 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
338E 27-Apr-2018 15:00:32 39 MRI_GXX 12 Gradient Power Amplifier error: Regulation error on X-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
339 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
340E 27-Apr-2018 15:00:32 21 MRI_GXX 11 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
341 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
342E 27-Apr-2018 15:00:32 2004 MRI_GXX 10 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
343 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
344E 27-Apr-2018 15:00:32 2000 MRI_GXX 9 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service
345 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
346E 27-Apr-2018 15:00:32 39 MRI_GXX 8 Gradient Power Amplifier error: Regulation error on X-axis. Check sequence parameter, retry. If the problem still persists, please call Siemens Service
347 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
348E 27-Apr-2018 15:00:32 21 MRI_GXX 7 Gradient Power Amplifier error: GPA has been switched off by the Stimulation-Monitor. Check sequence parameter, reboot Scanner. If the problem still persists, please call Siemens Service
349 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
350E 27-Apr-2018 15:00:32 2004 MRI_GXX 6 Gradient Power Amplifier error: Stimulation limit exceeded. Change the sequence parameter.
351 Time: 27.4.2018, 14:57:26, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
352E 27-Apr-2018 15:00:32 9 MRI_STD 5 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
353 Time: 27.4.2018, 14:57:22, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
354E 27-Apr-2018 15:00:32 801 MRI_COH 4 Scanner control error: Measurement is blocked by the MSUP supervisory test. Retry after the supervisory test is finished.
355 Time: 27.4.2018, 14:57:21, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
356E 27-Apr-2018 15:00:32 67 MRI_MSP 2 Magnet Supervision message: The MSU selftest is active, no scanning possible at this time. Wait for end of MSU selftest (approx. 1 minute). If the problem still persists, please call Siemens Service.
357 Time: 27.4.2018, 14:57:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
358E 27-Apr-2018 15:00:32 801 MRI_COH 1 Scanner control error: Measurement is blocked by the MSUP supervisory test. Retry after the supervisory test is finished.
359 Time: 27.4.2018, 14:57:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
360E 27-Apr-2018 14:58:56 27 MRI_MIF 0 Preparation of measurement system failed.
361 Time: 27.4.2018, 14:58:56, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
362 Text: MeasStatus not OK after AdjServerIF::adjust()
363E 27-Apr-2018 14:58:56 2010 MRI_ADS 0 Frequency adjustment did not converge. System could not be adjusted. Retry, reboot system, call service.
364 Time: 27.4.2018, 14:58:56, Line: 1586, File: \n4\pkg\MrServers\MrAdjustSrv\AdjServer\AdjControl\AdjControl.cpp, Process: CsaGenericMain (6232),
365 Text: Adjustment did not converge!
366E 27-Apr-2018 14:58:56 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service.
367 Time: 27.4.2018, 14:58:56, Line: 1585, File: \n4\pkg\MrServers\MrAdjustSrv\AdjServer\AdjControl\AdjControl.cpp, Process: CsaGenericMain (6232),
368 Text: Adjustment is invalid!
369E 27-Apr-2018 14:58:56 20 MRI_MSR 1 Preparation of Image Reconstruction System failed.
370 Time: 27.4.2018, 14:58:56, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
371 Text: CMD_PREP failed
372E 27-Apr-2018 14:58:56 38 MRI_MSR 0 Preparation of Image Reconstruction System failed.
373 Time: 27.4.2018, 14:58:56, Line: 3354, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
374 Text: Prepare of Acq failed
375E 27-Apr-2018 14:58:56 10 MRI_COS 1 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
376 Time: 27.4.2018, 14:58:56, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996),
377 Text: [LOG] Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
378E 27-Apr-2018 14:58:56 10 MRI_COS 0 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
379 Time: 27.4.2018, 14:58:10, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996)
380E 27-Apr-2018 14:58:10 27 MRI_MIF 0 Preparation of measurement system failed.
381 Time: 27.4.2018, 14:58:10, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
382 Text: MeasStatus not OK after AdjServerIF::adjust()
383E 27-Apr-2018 14:58:10 21 MRI_MSR 2 Preparation of Measurement System failed.
384 Time: 27.4.2018, 14:58:10, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
385 Text: CMD_PREP failed
386E 27-Apr-2018 14:58:10 38 MRI_MSR 1 Preparation of Image Reconstruction System failed.
387 Time: 27.4.2018, 14:58:10, Line: 3354, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
388 Text: Prepare of Acq failed
389E 27-Apr-2018 14:58:10 6 MRI_RFS 0 No patient registered.
390 Time: 27.4.2018, 14:57:58, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6232),
391 Text: ERROR: Table home but no Patient registered
392E 27-Apr-2018 14:58:10 10 MRI_COS 1 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
393 Time: 27.4.2018, 14:58:10, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996),
394 Text: [LOG] Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
395E 27-Apr-2018 14:58:10 10 MRI_COS 0 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
396 Time: 27.4.2018, 14:54:59, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996)
397E 27-Apr-2018 14:57:58 6 MRI_RFS 0 No patient registered.
398 Time: 27.4.2018, 14:57:58, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6504),
399 Text: ERROR: Table home but no Patient registered
400E 27-Apr-2018 14:56:09 34 MRI_PER 0 Scanner is booting.
401 Time: 27.4.2018, 14:56:09, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
402E 27-Apr-2018 14:55:22 35 MRI_PER 0 Scanner is not online (might be switched off).
403 Time: 27.4.2018, 14:55:22, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
404E 27-Apr-2018 14:55:22 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
405 Time: 27.4.2018, 14:55:22, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
406E 27-Apr-2018 14:54:59 27 MRI_MIF 0 Preparation of measurement system failed.
407 Time: 27.4.2018, 14:54:59, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
408 Text: MeasStatus not OK after AdjServerIF::adjust()
409E 27-Apr-2018 14:54:59 38 MRI_MSR 0 Preparation of Image Reconstruction System failed.
410 Time: 27.4.2018, 14:54:59, Line: 3354, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
411 Text: Prepare of Acq failed
412E 27-Apr-2018 14:54:59 10 MRI_COS 1 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
413 Time: 27.4.2018, 14:54:59, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996),
414 Text: [LOG] Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
415E 27-Apr-2018 14:54:59 10 MRI_COS 0 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
416 Time: 27.4.2018, 14:54:35, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996)
417E 27-Apr-2018 14:54:35 27 MRI_MIF 0 Preparation of measurement system failed.
418 Time: 27.4.2018, 14:54:35, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
419 Text: MeasStatus not OK after AdjServerIF::adjust()
420E 27-Apr-2018 14:54:35 21 MRI_MSR 1 Preparation of Measurement System failed.
421 Time: 27.4.2018, 14:54:35, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
422 Text: CMD_PREP failed
423E 27-Apr-2018 14:54:35 6 MRI_RFS 0 No patient registered.
424 Time: 27.4.2018, 14:54:20, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6232),
425 Text: ERROR: Table home but no Patient registered
426E 27-Apr-2018 14:54:35 10 MRI_COS 1 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
427 Time: 27.4.2018, 14:54:35, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996),
428 Text: [LOG] Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
429E 27-Apr-2018 14:54:35 10 MRI_COS 0 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
430 Time: 27.4.2018, 14:54:11, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996)
431E 27-Apr-2018 14:54:20 6 MRI_RFS 0 No patient registered.
432 Time: 27.4.2018, 14:54:20, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6504),
433 Text: ERROR: Table home but no Patient registered
434E 27-Apr-2018 14:54:11 6 MRI_RFS 0 No patient registered.
435 Time: 27.4.2018, 14:54:11, Process: CsaGenericMain (6232)
436E 27-Apr-2018 14:54:11 21 MRI_MSR 2 Preparation of Measurement System failed.
437 Time: 27.4.2018, 14:54:11, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
438 Text: CMD_PREP failed
439E 27-Apr-2018 14:54:11 38 MRI_MSR 1 Preparation of Image Reconstruction System failed.
440 Time: 27.4.2018, 14:54:11, Line: 3354, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
441 Text: Prepare of Acq failed
442E 27-Apr-2018 14:54:11 6 MRI_RFS 0 No patient registered.
443 Time: 27.4.2018, 14:53:55, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6232),
444 Text: ERROR: Table home but no Patient registered
445E 27-Apr-2018 14:54:11 10 MRI_COS 1 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
446 Time: 27.4.2018, 14:54:11, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996),
447 Text: [LOG] Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
448E 27-Apr-2018 14:54:11 10 MRI_COS 0 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
449 Time: 27.4.2018, 14:53:46, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996)
450E 27-Apr-2018 14:53:55 6 MRI_RFS 0 No patient registered.
451 Time: 27.4.2018, 14:53:55, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6504),
452 Text: ERROR: Table home but no Patient registered
453E 27-Apr-2018 14:53:46 20 MRI_MSR 0 Preparation of Image Reconstruction System failed.
454 Time: 27.4.2018, 14:53:46, Line: 128, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\IF\MesSerPrepareRet.cpp, Process: CsaGenericMain (6232),
455 Text: Received error status during MesSer prepare.
456E 27-Apr-2018 14:53:46 20 MRI_MSR 1 Preparation of Image Reconstruction System failed.
457 Time: 27.4.2018, 14:53:46, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
458 Text: CMD_PREP failed
459E 27-Apr-2018 14:53:46 38 MRI_MSR 0 Preparation of Image Reconstruction System failed.
460 Time: 27.4.2018, 14:53:46, Line: 3354, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
461 Text: Prepare of Acq failed
462E 27-Apr-2018 14:53:46 10 MRI_COS 1 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
463 Time: 27.4.2018, 14:53:46, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996),
464 Text: [LOG] Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
465E 27-Apr-2018 14:53:46 10 MRI_COS 0 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
466 Time: 27.4.2018, 14:53:22, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996)
467E 27-Apr-2018 14:53:22 6 MRI_RFS 0 No patient registered.
468 Time: 27.4.2018, 14:53:22, Process: CsaGenericMain (6232)
469E 27-Apr-2018 14:53:22 21 MRI_MSR 1 Preparation of Measurement System failed.
470 Time: 27.4.2018, 14:53:22, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
471 Text: CMD_PREP failed
472E 27-Apr-2018 14:53:22 6 MRI_RFS 0 No patient registered.
473 Time: 27.4.2018, 14:53:10, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6232),
474 Text: ERROR: Table home but no Patient registered
475E 27-Apr-2018 14:53:22 10 MRI_COS 1 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
476 Time: 27.4.2018, 14:53:22, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996),
477 Text: [LOG] Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
478E 27-Apr-2018 14:53:22 10 MRI_COS 0 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
479 Time: 27.4.2018, 14:52:54, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996)
480E 27-Apr-2018 14:53:10 6 MRI_RFS 0 No patient registered.
481 Time: 27.4.2018, 14:53:10, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6504),
482 Text: ERROR: Table home but no Patient registered
483E 27-Apr-2018 14:52:54 6 MRI_RFS 0 No patient registered.
484 Time: 27.4.2018, 14:52:54, Process: CsaGenericMain (6232)
485E 27-Apr-2018 14:52:54 21 MRI_MSR 1 Preparation of Measurement System failed.
486 Time: 27.4.2018, 14:52:54, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
487 Text: CMD_PREP failed
488E 27-Apr-2018 14:52:54 6 MRI_RFS 0 No patient registered.
489 Time: 27.4.2018, 14:52:35, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6232),
490 Text: ERROR: Table home but no Patient registered
491E 27-Apr-2018 14:52:54 10 MRI_COS 0 Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
492 Time: 27.4.2018, 14:52:54, Line: 3970, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\CocosIFImpl.cpp, Process: CsaGenericMain (4996),
493 Text: [LOG] Preparation of image reconstruction failed. System might be busy. Please retry measurement, if not successful restart syngo MR.
494E 27-Apr-2018 14:52:35 8 MRI_PRC 0 Image reconstruction failed. MeasUid 303.
495 Time: 27.4.2018, 14:51:01, Line: 62, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\DispatcherBase.cpp, Process: CsaGenericMain (4996),
496 Text: SPICE
497E 27-Apr-2018 14:52:35 6 MRI_RFS 0 No patient registered.
498 Time: 27.4.2018, 14:52:35, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6504),
499 Text: ERROR: Table home but no Patient registered
500E 27-Apr-2018 14:51:01 8 MRI_PRC 1 Image reconstruction failed. MeasUid 303.
501 Time: 27.4.2018, 14:51:01, Line: 62, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\DispatcherBase.cpp, Process: CsaGenericMain (4996),
502 Text: SPICE
503 [LOG] Image reconstruction failed. MeasUid 303.
504E 27-Apr-2018 14:48:29 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service.
505 Time: 27.4.2018, 14:48:29, Line: 1036, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
506 Text: SPRINT error
507E 27-Apr-2018 14:48:29 12 MRI_MSR 1 Scanning stopped.
508 Time: 27.4.2018, 14:48:29, Line: 105, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\IF\MesSerStartRet.cpp, Process: CsaGenericMain (6232),
509 Text: Received error status during MesSer start.
510E 27-Apr-2018 14:48:29 12 MRI_MSR 0 Scanning stopped.
511 Time: 27.4.2018, 14:48:29, Line: 105, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\IF\MesSerStartRet.cpp, Process: CsaGenericMain (6232),
512 Text: Received error status during MesSer start.
513E 27-Apr-2018 14:48:29 12 MRI_MSR 4 Scanning stopped.
514 Time: 27.4.2018, 14:48:29, Line: 4463, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
515 Text: measurement failed
516E 27-Apr-2018 14:48:29 25 MRI_MSR 0 Scanning stopped by Image Reconstruction.
517 Time: 27.4.2018, 14:48:29, Line: 4257, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
518 Text: Data acquiasition finished with error.
519E 27-Apr-2018 14:48:26 32 MRI_COS 0 Connection to Image Reconstruction System has been lost.
520 Time: 27.4.2018, 14:48:26, Line: 264, File: \n4\pkg\MrServers\MrVista\Cocos\MrCocosIPCProxyComp\MrIrisPCClientWatchdog.cpp, Process: CsaGenericMain (5116),
521 Text: [LOG] Connection to Image Reconstruction System has been lost.
522E 27-Apr-2018 14:14:55 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
523E 27-Apr-2018 14:13:55 27 MRI_MIF 0 Preparation of measurement system failed.
524 Time: 27.4.2018, 14:13:55, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
525 Text: MeasStatus not OK after AdjServerIF::adjust()
526E 27-Apr-2018 14:13:55 21 MRI_MSR 1 Preparation of Measurement System failed.
527 Time: 27.4.2018, 14:13:55, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
528 Text: CMD_PREP failed
529E 27-Apr-2018 14:13:55 7 MRI_RFS 0 No patient positioned.
530 Time: 27.4.2018, 14:13:54, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6232),
531 Text: ERROR: Patient registered but no table home
532E 27-Apr-2018 14:13:54 7 MRI_RFS 0 No patient positioned.
533 Time: 27.4.2018, 14:13:54, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6504),
534 Text: ERROR: Patient registered but no table home
535E 27-Apr-2018 13:36:38 27 MRI_MIF 0 Preparation of measurement system failed.
536 Time: 27.4.2018, 13:36:38, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
537 Text: MeasStatus not OK after AdjServerIF::adjust()
538E 27-Apr-2018 13:36:38 21 MRI_MSR 1 Preparation of Measurement System failed.
539 Time: 27.4.2018, 13:36:38, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
540 Text: CMD_PREP failed
541E 27-Apr-2018 13:36:38 7 MRI_RFS 0 No patient positioned.
542 Time: 27.4.2018, 13:36:37, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6232),
543 Text: ERROR: Patient registered but no table home
544E 27-Apr-2018 13:36:37 7 MRI_RFS 0 No patient positioned.
545 Time: 27.4.2018, 13:36:37, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6504),
546 Text: ERROR: Patient registered but no table home
547E 27-Apr-2018 12:34:49 2 MRI_ARG 0 No images could be loaded. Either images selected were duplicates of those already under study, or the type of image data selected could not be loaded in Argus.
548 Time: 27.4.2018, 12:34:49, Line: 356, File: \n4\pkg\MrApplications\MrArgus\FrontEnd\ProtocolFE\arguscvprotocolfe\arguscvprotocolfectl.cpp, Process: CapGM (7848)
549E 27-Apr-2018 12:34:49 2 MRI_ARG 0 No images could be loaded. Either images selected were duplicates of those already under study, or the type of image data selected could not be loaded in Argus.
550 Time: 27.4.2018, 12:34:49, Line: 535, File: \n4\pkg\MrApplications\MrArgus\FrontEnd\ProtocolFE\arguscvprotocolfe\arguscvprotocolfectl.cpp, Process: CapGM (7848)
551E 27-Apr-2018 12:30:30 27 MRI_MIF 0 Preparation of measurement system failed.
552 Time: 27.4.2018, 12:30:30, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
553 Text: MeasStatus not OK after AdjServerIF::adjust()
554E 27-Apr-2018 12:30:30 21 MRI_MSR 1 Preparation of Measurement System failed.
555 Time: 27.4.2018, 12:30:30, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
556 Text: CMD_PREP failed
557E 27-Apr-2018 12:30:30 7 MRI_RFS 0 No patient positioned.
558 Time: 27.4.2018, 12:30:30, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6232),
559 Text: ERROR: Patient registered but no table home
560E 27-Apr-2018 12:30:30 7 MRI_RFS 0 No patient positioned.
561 Time: 27.4.2018, 12:30:30, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6504),
562 Text: ERROR: Patient registered but no table home
563E 27-Apr-2018 12:06:32 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
564E 27-Apr-2018 11:56:15 7 MRI_MIF 1 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service.
565 Time: 27.4.2018, 11:56:15, Line: 1036, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
566 Text: SPRINT error
567E 27-Apr-2018 11:55:28 8 MRI_RFS 0 SAR limit(s) exceeded.
568 Time: 27.4.2018, 11:55:28, Process: CsaGenericMain (6232)
569E 27-Apr-2018 11:55:28 21 MRI_MSR 1 Preparation of Measurement System failed.
570 Time: 27.4.2018, 11:55:28, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
571 Text: CMD_PREP failed
572E 27-Apr-2018 11:55:28 8 MRI_RFS 0 SAR limit(s) exceeded.
573 Time: 27.4.2018, 11:55:28, Line: 2344, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6232),
574 Text: ERROR: Prediction for SAR-Values: Limits exceeded
575E 27-Apr-2018 11:55:28 8 MRI_RFS 0 SAR limit(s) exceeded.
576 Time: 27.4.2018, 11:55:28, Line: 2344, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6504),
577 Text: ERROR: Prediction for SAR-Values: Limits exceeded
578E 27-Apr-2018 11:41:28 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
579E 27-Apr-2018 11:35:59 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
580E 27-Apr-2018 09:10:52 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
581E 27-Apr-2018 08:40:08 7 MRI_MIF 1 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service.
582 Time: 27.4.2018, 08:40:08, Line: 1036, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
583 Text: SPRINT error
584E 27-Apr-2018 08:39:34 27 MRI_MIF 0 Preparation of measurement system failed.
585 Time: 27.4.2018, 08:39:34, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
586 Text: MeasStatus not OK after AdjServerIF::adjust()
587E 27-Apr-2018 08:39:34 21 MRI_MSR 1 Preparation of Measurement System failed.
588 Time: 27.4.2018, 08:39:34, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6232),
589 Text: CMD_PREP failed
590E 27-Apr-2018 08:39:34 6 MRI_RFS 0 No patient registered.
591 Time: 27.4.2018, 08:39:33, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6232),
592 Text: ERROR: Table home but no Patient registered
593E 27-Apr-2018 08:39:33 6 MRI_RFS 0 No patient registered.
594 Time: 27.4.2018, 08:39:33, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6504),
595 Text: ERROR: Table home but no Patient registered
596E 27-Apr-2018 08:19:45 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
597E 27-Apr-2018 08:16:35 7 MRI_MIF 1 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service.
598 Time: 27.4.2018, 08:16:35, Line: 1036, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6232),
599 Text: SPRINT error
600E 27-Apr-2018 07:40:59 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
601E 27-Apr-2018 07:40:59 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
602E 27-Apr-2018 07:39:39 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
603E 27-Apr-2018 07:20:15 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
604E 27-Apr-2018 07:19:37 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
605E 27-Apr-2018 07:19:06 1 CAP_DVF 0 An unspecified internal error has occurred.
606 Time: 27.4.2018, 07:19:06, Line: 271, File: \MedCom\src\3D\Visualization\FrontEnd\Editor\src/Cap3dFEEditorDlgInit.cpp, Process: CapGM (7848),
607 Text: "It exists no valid Editor license."
608E 27-Apr-2018 07:18:29 34 MRI_PER 0 Scanner is booting.
609 Time: 27.4.2018, 07:18:29, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
610E 27-Apr-2018 07:18:26 35 MRI_PER 0 Scanner is not online (might be switched off).
611 Time: 27.4.2018, 07:18:26, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6644)
612E 27-Apr-2018 07:18:26 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
613 Time: 27.4.2018, 07:18:25, Process: CsaGenericMain (6504)
614E 27-Apr-2018 07:18:18 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection.
615 Time: 27.4.2018, 07:18:18, Process: CsaGenericMain (6232)
616E 27-Apr-2018 07:18:17 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
617 Time: 27.4.2018, 07:18:17, Process: CsaGenericMain (6232)
618E 27-Apr-2018 07:17:16 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
619E 27-Apr-2018 07:17:10 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
620E 27-Apr-2018 07:17:10 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
621E 27-Apr-2018 07:17:09 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
622E 27-Apr-2018 07:17:08 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
623E 27-Apr-2018 07:17:06 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
624E 27-Apr-2018 07:17:06 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
625E 27-Apr-2018 07:17:06 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
626E 27-Apr-2018 07:17:05 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
627E 27-Apr-2018 07:17:05 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
628E 27-Apr-2018 07:17:04 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
629E 27-Apr-2018 07:16:57 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
630E 27-Apr-2018 07:16:56 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
631E 27-Apr-2018 07:16:55 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
632E 27-Apr-2018 07:16:55 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
633E 27-Apr-2018 07:16:48 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
634E 27-Apr-2018 07:16:44 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
635E 27-Apr-2018 07:16:17 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Services\TmFilter\Parameters' with value 'Cue' by process C:\Program Files (x86)\Trend Micro\OfficeScan Client\NTRTScan.exe (Process Id: 2192, User: NT AUTHORITY\SYSTEM).
636E 27-Apr-2018 07:16:16 20 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify file 'c:\program files\mcafee\solidcore\sadmin.exe' by process C:\WINNT\system32\csrss.exe (Process Id: 464, User: NT AUTHORITY\SYSTEM).
637E 27-Apr-2018 07:16:16 20 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify file 'c:\program files\mcafee\solidcore\sadmin.exe' by process C:\WINNT\system32\csrss.exe (Process Id: 464, User: NT AUTHORITY\SYSTEM).
638E 27-Apr-2018 07:16:06 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'ctfmon.exe' by process C:\WINNT\SysWOW64\ctfmon.exe (Process Id: 1880, User: MRC31210\meduser).
639E 27-Apr-2018 07:16:05 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'internat.exe' by process C:\WINNT\system32\ctfmon.exe (Process Id: 1756, User: MRC31210\meduser).
640E 27-Apr-2018 07:16:05 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'ctfmon.exe' by process C:\WINNT\system32\ctfmon.exe (Process Id: 1756, User: MRC31210\meduser).
641E 27-Apr-2018 07:15:57 50 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_LOCAL_MACHINE\SOFTWARE\Classes\Wow6432Node' by process C:\WINNT\system32\csrss.exe (Process Id: 464, User: NT AUTHORITY\SYSTEM).
642E 26-Apr-2018 21:44:52 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
643E 26-Apr-2018 21:44:51 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
644E 26-Apr-2018 21:44:51 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
645E 26-Apr-2018 21:44:50 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
646E 26-Apr-2018 21:44:02 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
647E 26-Apr-2018 21:43:50 32 MRI_COS 6 Connection to Image Reconstruction System has been lost.
648 Time: 26.4.2018, 21:41:46, Line: 876, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4708)
649E 26-Apr-2018 21:43:50 11 MRI_COS 5 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords.
650 Time: 26.4.2018, 21:41:38, Line: 838, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4708)
651E 26-Apr-2018 21:43:50 11 MRI_COS 3 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords.
652 Time: 26.4.2018, 17:27:48, Line: 793, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4708)
653E 26-Apr-2018 21:43:50 11 MRI_COS 2 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords.
654 Time: 26.4.2018, 17:27:47, Line: 838, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4708)
655E 26-Apr-2018 21:43:50 32 MRI_COS 1 Connection to Image Reconstruction System has been lost.
656 Time: 26.4.2018, 17:27:47, Line: 862, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4708)
657E 26-Apr-2018 21:43:50 32 MRI_COS 0 Connection to Image Reconstruction System has been lost.
658 Time: 26.4.2018, 17:27:41, Line: 876, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4708)
659E 26-Apr-2018 21:43:39 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details.
660 Time: 26.4.2018, 21:43:39, Line: 156, File: \n4\pkg\MrApplications\MrExamUI\MrOnlineDisplay\MrDisplayerSvc\MrDisplayerDBSession.cpp, Process: CsaGenericMain (4140),
661 Text: Failed to end session
662 [LOG] An error occurred during accessing db. See logfile for details.
663E 26-Apr-2018 21:43:39 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
664E 26-Apr-2018 21:43:37 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
665 Time: 26.4.2018, 21:42:24, Process: CsaGenericMain (5288)
666E 26-Apr-2018 21:42:58 35 MRI_PER 29 Scanner is not online (might be switched off).
667 Time: 26.4.2018, 21:42:24, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6624)
668E 26-Apr-2018 21:42:58 1 MRI_PER 28 Scanner is not online. Wait until booting has finished, or switch on Scanner.
669 Time: 26.4.2018, 21:42:24, Process: CsaGenericMain (6624)
670E 26-Apr-2018 21:42:58 15 MRI_PER 27 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
671 Time: 26.4.2018, 21:42:24, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6624)
672E 26-Apr-2018 21:42:58 2009 MRI_PXX 26 Patient table error: Voltage error (24V DC ORT). Please call Siemens Service.
673 Time: 26.4.2018, 21:42:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6624)
674E 26-Apr-2018 21:42:58 2008 MRI_PXX 25 Patient table error: Voltage error (24V DC SWT). Please call Siemens Service.
675 Time: 26.4.2018, 21:42:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6624)
676E 26-Apr-2018 21:42:58 9 MRI_STD 24 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
677 Time: 26.4.2018, 21:42:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6624)
678E 26-Apr-2018 21:42:58 207 MRI_SUP 23 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
679 Time: 26.4.2018, 21:42:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6624)
680E 26-Apr-2018 21:42:58 32 MRI_COH 22 Patient table error: Table position check failed Please retry the measurement. Please save System Log-Files immediately by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service.
681 Time: 26.4.2018, 21:37:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6624)
682E 26-Apr-2018 21:42:58 3000 MRI_PXX 16 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service
683 Time: 26.4.2018, 21:28:50, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6624)
684E 26-Apr-2018 21:42:58 440 MRI_SUP 15 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service.
685 Time: 26.4.2018, 21:28:50, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6624)
686E 26-Apr-2018 21:42:58 430 MRI_SUP 14 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service.
687 Time: 26.4.2018, 21:28:50, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6624)
688E 26-Apr-2018 21:42:48 35 MRI_PER 30 Scanner is not online (might be switched off).
689 Time: 26.4.2018, 21:42:24, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
690E 26-Apr-2018 21:42:48 15 MRI_PER 29 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
691 Time: 26.4.2018, 21:42:24, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
692E 26-Apr-2018 21:42:48 2009 MRI_PXX 28 Patient table error: Voltage error (24V DC ORT). Please call Siemens Service.
693 Time: 26.4.2018, 21:42:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
694E 26-Apr-2018 21:42:48 2008 MRI_PXX 27 Patient table error: Voltage error (24V DC SWT). Please call Siemens Service.
695 Time: 26.4.2018, 21:42:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
696E 26-Apr-2018 21:42:48 9 MRI_STD 26 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
697 Time: 26.4.2018, 21:42:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
698E 26-Apr-2018 21:42:48 207 MRI_SUP 25 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
699 Time: 26.4.2018, 21:42:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
700E 26-Apr-2018 21:42:48 32 MRI_COH 24 Patient table error: Table position check failed Please retry the measurement. Please save System Log-Files immediately by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service.
701 Time: 26.4.2018, 21:37:11, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
702E 26-Apr-2018 21:42:48 3000 MRI_PXX 18 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service
703 Time: 26.4.2018, 21:28:50, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
704E 26-Apr-2018 21:42:48 440 MRI_SUP 17 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service.
705 Time: 26.4.2018, 21:28:50, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
706E 26-Apr-2018 21:42:48 430 MRI_SUP 16 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service.
707 Time: 26.4.2018, 21:28:50, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
708E 26-Apr-2018 21:42:24 35 MRI_PER 0 Scanner is not online (might be switched off).
709 Time: 26.4.2018, 21:42:24, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
710E 26-Apr-2018 21:42:24 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
711 Time: 26.4.2018, 21:42:24, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
712E 26-Apr-2018 21:42:11 2009 MRI_PXX 0 Patient table error: Voltage error (24V DC ORT). Please call Siemens Service.
713 Time: 26.4.2018, 21:42:11, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
714E 26-Apr-2018 21:42:11 2008 MRI_PXX 0 Patient table error: Voltage error (24V DC SWT). Please call Siemens Service.
715 Time: 26.4.2018, 21:42:11, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
716E 26-Apr-2018 21:42:11 9 MRI_STD 0 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
717 Time: 26.4.2018, 21:42:11, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
718E 26-Apr-2018 21:42:11 207 MRI_SUP 0 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
719 Time: 26.4.2018, 21:42:11, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
720E 26-Apr-2018 18:40:07 27 MRI_MIF 0 Preparation of measurement system failed.
721 Time: 26.4.2018, 18:40:07, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6260),
722 Text: MeasStatus not OK after AdjServerIF::adjust()
723E 26-Apr-2018 18:40:07 21 MRI_MSR 1 Preparation of Measurement System failed.
724 Time: 26.4.2018, 18:40:07, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6260),
725 Text: CMD_PREP failed
726E 26-Apr-2018 18:40:07 6 MRI_RFS 0 No patient registered.
727 Time: 26.4.2018, 18:40:07, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6260),
728 Text: ERROR: Table home but no Patient registered
729E 26-Apr-2018 18:40:07 6 MRI_RFS 0 No patient registered.
730 Time: 26.4.2018, 18:40:07, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6532),
731 Text: ERROR: Table home but no Patient registered
732E 26-Apr-2018 18:38:23 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
733E 26-Apr-2018 18:38:23 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
734E 26-Apr-2018 17:30:41 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
735E 26-Apr-2018 17:30:34 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
736E 26-Apr-2018 17:30:14 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
737E 26-Apr-2018 17:29:56 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
738E 26-Apr-2018 17:29:24 1 CAP_DVF 0 An unspecified internal error has occurred.
739 Time: 26.4.2018, 17:29:24, Line: 271, File: \MedCom\src\3D\Visualization\FrontEnd\Editor\src/Cap3dFEEditorDlgInit.cpp, Process: CapGM (7836),
740 Text: "It exists no valid Editor license."
741E 26-Apr-2018 17:28:47 34 MRI_PER 0 Scanner is booting.
742 Time: 26.4.2018, 17:28:47, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
743E 26-Apr-2018 17:28:44 35 MRI_PER 0 Scanner is not online (might be switched off).
744 Time: 26.4.2018, 17:28:44, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6676)
745E 26-Apr-2018 17:28:44 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
746 Time: 26.4.2018, 17:28:43, Process: CsaGenericMain (6532)
747E 26-Apr-2018 17:28:36 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection.
748 Time: 26.4.2018, 17:28:36, Process: CsaGenericMain (6260)
749E 26-Apr-2018 17:28:35 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
750 Time: 26.4.2018, 17:28:35, Process: CsaGenericMain (6260)
751E 26-Apr-2018 17:27:36 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
752E 26-Apr-2018 17:27:30 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
753E 26-Apr-2018 17:27:30 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
754E 26-Apr-2018 17:27:29 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
755E 26-Apr-2018 17:27:29 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
756E 26-Apr-2018 17:27:27 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
757E 26-Apr-2018 17:27:27 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
758E 26-Apr-2018 17:27:26 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
759E 26-Apr-2018 17:27:25 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
760E 26-Apr-2018 17:27:25 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
761E 26-Apr-2018 17:27:24 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
762E 26-Apr-2018 17:27:17 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
763E 26-Apr-2018 17:27:16 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
764E 26-Apr-2018 17:27:16 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
765E 26-Apr-2018 17:27:15 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
766E 26-Apr-2018 17:27:08 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
767E 26-Apr-2018 17:27:06 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
768E 26-Apr-2018 17:26:41 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Services\TmFilter\Parameters' with value 'Cue' by process C:\Program Files (x86)\Trend Micro\OfficeScan Client\NTRTScan.exe (Process Id: 2148, User: NT AUTHORITY\SYSTEM).
769E 26-Apr-2018 17:26:40 20 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify file 'c:\program files\mcafee\solidcore\sadmin.exe' by process C:\WINNT\system32\csrss.exe (Process Id: 468, User: NT AUTHORITY\SYSTEM).
770E 26-Apr-2018 17:26:39 20 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify file 'c:\program files\mcafee\solidcore\sadmin.exe' by process C:\WINNT\system32\csrss.exe (Process Id: 468, User: NT AUTHORITY\SYSTEM).
771E 26-Apr-2018 17:26:30 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'ctfmon.exe' by process C:\WINNT\SysWOW64\ctfmon.exe (Process Id: 1868, User: MRC31210\meduser).
772E 26-Apr-2018 17:26:29 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'internat.exe' by process C:\WINNT\system32\ctfmon.exe (Process Id: 1764, User: MRC31210\meduser).
773E 26-Apr-2018 17:26:29 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'ctfmon.exe' by process C:\WINNT\system32\ctfmon.exe (Process Id: 1764, User: MRC31210\meduser).
774E 26-Apr-2018 17:26:21 50 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_LOCAL_MACHINE\SOFTWARE\Classes\Wow6432Node' by process C:\WINNT\system32\csrss.exe (Process Id: 468, User: NT AUTHORITY\SYSTEM).
775E 26-Apr-2018 17:21:44 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
776E 26-Apr-2018 17:21:43 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
777E 26-Apr-2018 17:21:42 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
778E 26-Apr-2018 17:21:42 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
779E 26-Apr-2018 17:20:45 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
780E 26-Apr-2018 17:20:32 32 MRI_COS 3 Connection to Image Reconstruction System has been lost.
781 Time: 26.4.2018, 17:14:20, Line: 876, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (5008)
782E 26-Apr-2018 17:20:32 11 MRI_COS 1 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords.
783 Time: 26.4.2018, 07:12:07, Line: 793, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (5008)
784E 26-Apr-2018 17:20:32 32 MRI_COS 0 Connection to Image Reconstruction System has been lost.
785 Time: 26.4.2018, 07:12:07, Line: 862, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (5008)
786E 26-Apr-2018 17:20:21 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details.
787 Time: 26.4.2018, 17:20:21, Line: 156, File: \n4\pkg\MrApplications\MrExamUI\MrOnlineDisplay\MrDisplayerSvc\MrDisplayerDBSession.cpp, Process: CsaGenericMain (4260),
788 Text: Failed to end session
789 [LOG] An error occurred during accessing db. See logfile for details.
790E 26-Apr-2018 17:20:21 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
791E 26-Apr-2018 17:20:20 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
792 Time: 26.4.2018, 17:19:06, Process: CsaGenericMain (5176)
793E 26-Apr-2018 17:19:40 35 MRI_PER 29 Scanner is not online (might be switched off).
794 Time: 26.4.2018, 17:19:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6992)
795E 26-Apr-2018 17:19:40 1 MRI_PER 28 Scanner is not online. Wait until booting has finished, or switch on Scanner.
796 Time: 26.4.2018, 17:19:06, Process: CsaGenericMain (6992)
797E 26-Apr-2018 17:19:40 15 MRI_PER 27 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
798 Time: 26.4.2018, 17:19:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6992)
799E 26-Apr-2018 17:19:40 28 MRI_SLO 26 RF Power Amplifier error: Temperature of the RFPA too high. Retry after 5 minutes. If the problem still persists, please call Siemens Service.
800 Time: 26.4.2018, 17:18:52, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6992)
801E 26-Apr-2018 17:19:40 27 MRI_SLO 25 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
802 Time: 26.4.2018, 17:18:52, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6992)
803E 26-Apr-2018 17:19:40 9 MRI_STD 24 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
804 Time: 26.4.2018, 17:18:52, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6992)
805E 26-Apr-2018 17:19:40 207 MRI_SUP 23 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
806 Time: 26.4.2018, 17:18:52, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6992)
807E 26-Apr-2018 17:19:40 32 MRI_COH 22 Patient table error: Table position check failed Please retry the measurement. Please save System Log-Files immediately by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service.
808 Time: 26.4.2018, 17:15:28, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6992)
809E 26-Apr-2018 17:19:40 3000 MRI_PXX 8 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service
810 Time: 26.4.2018, 17:06:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6992)
811E 26-Apr-2018 17:19:40 440 MRI_SUP 7 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service.
812 Time: 26.4.2018, 17:06:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6992)
813E 26-Apr-2018 17:19:40 430 MRI_SUP 6 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service.
814 Time: 26.4.2018, 17:06:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6992)
815E 26-Apr-2018 17:19:30 35 MRI_PER 30 Scanner is not online (might be switched off).
816 Time: 26.4.2018, 17:19:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
817E 26-Apr-2018 17:19:30 15 MRI_PER 29 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
818 Time: 26.4.2018, 17:19:06, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
819E 26-Apr-2018 17:19:30 28 MRI_SLO 28 RF Power Amplifier error: Temperature of the RFPA too high. Retry after 5 minutes. If the problem still persists, please call Siemens Service.
820 Time: 26.4.2018, 17:18:52, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
821E 26-Apr-2018 17:19:30 27 MRI_SLO 27 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
822 Time: 26.4.2018, 17:18:52, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
823E 26-Apr-2018 17:19:30 9 MRI_STD 26 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
824 Time: 26.4.2018, 17:18:52, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
825E 26-Apr-2018 17:19:30 207 MRI_SUP 25 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
826 Time: 26.4.2018, 17:18:52, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
827E 26-Apr-2018 17:19:30 32 MRI_COH 24 Patient table error: Table position check failed Please retry the measurement. Please save System Log-Files immediately by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service.
828 Time: 26.4.2018, 17:15:28, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
829E 26-Apr-2018 17:19:30 3000 MRI_PXX 10 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service
830 Time: 26.4.2018, 17:06:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
831E 26-Apr-2018 17:19:30 440 MRI_SUP 9 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service.
832 Time: 26.4.2018, 17:06:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
833E 26-Apr-2018 17:19:30 430 MRI_SUP 8 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service.
834 Time: 26.4.2018, 17:06:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
835E 26-Apr-2018 17:19:06 35 MRI_PER 0 Scanner is not online (might be switched off).
836 Time: 26.4.2018, 17:19:06, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
837E 26-Apr-2018 17:19:06 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
838 Time: 26.4.2018, 17:19:06, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
839E 26-Apr-2018 17:18:52 9 MRI_STD 0 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
840 Time: 26.4.2018, 17:18:52, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
841E 26-Apr-2018 17:18:52 207 MRI_SUP 0 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
842 Time: 26.4.2018, 17:18:52, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
843E 26-Apr-2018 17:16:51 8 MRI_PRC 1 Image reconstruction failed. MeasUid 414.
844 Time: 26.4.2018, 17:16:51, Line: 62, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\DispatcherBase.cpp, Process: CsaGenericMain (5008),
845 Text: SPICE
846 [LOG] Image reconstruction failed. MeasUid 414.
847E 26-Apr-2018 17:14:20 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service.
848 Time: 26.4.2018, 17:14:20, Line: 1036, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6632),
849 Text: SPRINT error
850E 26-Apr-2018 17:14:20 12 MRI_MSR 1 Scanning stopped.
851 Time: 26.4.2018, 17:14:20, Line: 105, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\IF\MesSerStartRet.cpp, Process: CsaGenericMain (6632),
852 Text: Received error status during MesSer start.
853E 26-Apr-2018 17:14:20 12 MRI_MSR 0 Scanning stopped.
854 Time: 26.4.2018, 17:14:20, Line: 105, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\IF\MesSerStartRet.cpp, Process: CsaGenericMain (6632),
855 Text: Received error status during MesSer start.
856E 26-Apr-2018 17:14:20 12 MRI_MSR 4 Scanning stopped.
857 Time: 26.4.2018, 17:14:20, Line: 4463, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6632),
858 Text: measurement failed
859E 26-Apr-2018 17:14:20 25 MRI_MSR 0 Scanning stopped by Image Reconstruction.
860 Time: 26.4.2018, 17:14:20, Line: 4257, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6632),
861 Text: Data acquiasition finished with error.
862E 26-Apr-2018 17:14:16 32 MRI_COS 0 Connection to Image Reconstruction System has been lost.
863 Time: 26.4.2018, 17:14:16, Line: 264, File: \n4\pkg\MrServers\MrVista\Cocos\MrCocosIPCProxyComp\MrIrisPCClientWatchdog.cpp, Process: CsaGenericMain (5076),
864 Text: [LOG] Connection to Image Reconstruction System has been lost.
865E 26-Apr-2018 15:23:13 27 MRI_MIF 0 Preparation of measurement system failed.
866 Time: 26.4.2018, 15:23:13, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6632),
867 Text: MeasStatus not OK after AdjServerIF::adjust()
868E 26-Apr-2018 15:23:13 21 MRI_MSR 1 Preparation of Measurement System failed.
869 Time: 26.4.2018, 15:23:13, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6632),
870 Text: CMD_PREP failed
871E 26-Apr-2018 15:23:13 7 MRI_RFS 0 No patient positioned.
872 Time: 26.4.2018, 15:23:13, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6632),
873 Text: ERROR: Patient registered but no table home
874E 26-Apr-2018 15:23:13 7 MRI_RFS 0 No patient positioned.
875 Time: 26.4.2018, 15:23:13, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6904),
876 Text: ERROR: Patient registered but no table home
877E 26-Apr-2018 15:23:08 29 MRI_MSR 0 Invalid coil selection. Open protocol or check plugging of coils.
878 Time: 26.4.2018, 15:23:08, Line: 3189, File: \n4\pkg\MrServers\MrProtSrv\EdbProtPara\MrEdbProtPara.cpp, Process: CsaGenericMain (5372)
879E 26-Apr-2018 15:04:34 7 MRI_MIF 1 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service.
880 Time: 26.4.2018, 15:04:34, Line: 1036, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6632),
881 Text: SPRINT error
882E 26-Apr-2018 15:02:34 27 MRI_MIF 0 Preparation of measurement system failed.
883 Time: 26.4.2018, 15:02:34, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6632),
884 Text: MeasStatus not OK after AdjServerIF::adjust()
885E 26-Apr-2018 15:02:34 21 MRI_MSR 1 Preparation of Measurement System failed.
886 Time: 26.4.2018, 15:02:34, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6632),
887 Text: CMD_PREP failed
888E 26-Apr-2018 15:02:34 7 MRI_RFS 0 No patient positioned.
889 Time: 26.4.2018, 15:02:34, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6632),
890 Text: ERROR: Patient registered but no table home
891E 26-Apr-2018 15:02:34 7 MRI_RFS 0 No patient positioned.
892 Time: 26.4.2018, 15:02:34, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6904),
893 Text: ERROR: Patient registered but no table home
894E 26-Apr-2018 15:02:29 29 MRI_MSR 0 Invalid coil selection. Open protocol or check plugging of coils.
895 Time: 26.4.2018, 15:02:29, Line: 3189, File: \n4\pkg\MrServers\MrProtSrv\EdbProtPara\MrEdbProtPara.cpp, Process: CsaGenericMain (5372)
896E 26-Apr-2018 15:02:08 7 MRI_MIF 1 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service.
897 Time: 26.4.2018, 15:02:08, Line: 1036, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6632),
898 Text: SPRINT error
899E 26-Apr-2018 15:01:25 29 MRI_MSR 0 Invalid coil selection. Open protocol or check plugging of coils.
900 Time: 26.4.2018, 15:01:25, Line: 3189, File: \n4\pkg\MrServers\MrProtSrv\EdbProtPara\MrEdbProtPara.cpp, Process: CsaGenericMain (5372)
901E 26-Apr-2018 11:32:09 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
902E 26-Apr-2018 10:56:12 27 MRI_MIF 0 Preparation of measurement system failed.
903 Time: 26.4.2018, 10:56:12, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6632),
904 Text: MeasStatus not OK after AdjServerIF::adjust()
905E 26-Apr-2018 10:56:12 21 MRI_MSR 1 Preparation of Measurement System failed.
906 Time: 26.4.2018, 10:56:12, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6632),
907 Text: CMD_PREP failed
908E 26-Apr-2018 10:56:12 6 MRI_RFS 0 No patient registered.
909 Time: 26.4.2018, 10:56:12, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6632),
910 Text: ERROR: Table home but no Patient registered
911E 26-Apr-2018 10:56:12 6 MRI_RFS 0 No patient registered.
912 Time: 26.4.2018, 10:56:12, Line: 1854, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6904),
913 Text: ERROR: Table home but no Patient registered
914E 26-Apr-2018 10:01:56 8 MRI_RFS 0 SAR limit(s) exceeded.
915 Time: 26.4.2018, 10:01:56, Process: CsaGenericMain (6632)
916E 26-Apr-2018 10:01:56 21 MRI_MSR 1 Preparation of Measurement System failed.
917 Time: 26.4.2018, 10:01:56, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6632),
918 Text: CMD_PREP failed
919E 26-Apr-2018 10:01:56 8 MRI_RFS 0 SAR limit(s) exceeded.
920 Time: 26.4.2018, 10:01:55, Line: 2344, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6632),
921 Text: ERROR: Prediction for SAR-Values: Limits exceeded
922E 26-Apr-2018 10:01:55 8 MRI_RFS 0 SAR limit(s) exceeded.
923 Time: 26.4.2018, 10:01:55, Line: 2344, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6904),
924 Text: ERROR: Prediction for SAR-Values: Limits exceeded
925E 26-Apr-2018 08:57:05 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
926E 26-Apr-2018 08:20:07 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
927E 26-Apr-2018 07:48:37 7 MRI_MIF 1 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service.
928 Time: 26.4.2018, 07:48:37, Line: 1036, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6632),
929 Text: SPRINT error
930E 26-Apr-2018 07:48:29 27 MRI_MIF 0 Preparation of measurement system failed.
931 Time: 26.4.2018, 07:48:29, Line: 614, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6632),
932 Text: MeasStatus not OK after AdjServerIF::adjust()
933E 26-Apr-2018 07:48:29 21 MRI_MSR 1 Preparation of Measurement System failed.
934 Time: 26.4.2018, 07:48:29, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6632),
935 Text: CMD_PREP failed
936E 26-Apr-2018 07:48:29 7 MRI_RFS 0 No patient positioned.
937 Time: 26.4.2018, 07:48:28, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6632),
938 Text: ERROR: Patient registered but no table home
939E 26-Apr-2018 07:48:28 7 MRI_RFS 0 No patient positioned.
940 Time: 26.4.2018, 07:48:28, Line: 1860, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6904),
941 Text: ERROR: Patient registered but no table home
942E 26-Apr-2018 07:36:03 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
943E 26-Apr-2018 07:35:06 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
944E 26-Apr-2018 07:31:46 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
945E 26-Apr-2018 07:15:01 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
946E 26-Apr-2018 07:14:25 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
947E 26-Apr-2018 07:13:51 1 CAP_DVF 0 An unspecified internal error has occurred.
948 Time: 26.4.2018, 07:13:51, Line: 271, File: \MedCom\src\3D\Visualization\FrontEnd\Editor\src/Cap3dFEEditorDlgInit.cpp, Process: CapGM (7880),
949 Text: "It exists no valid Editor license."
950E 26-Apr-2018 07:13:13 34 MRI_PER 0 Scanner is booting.
951 Time: 26.4.2018, 07:13:13, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
952E 26-Apr-2018 07:13:10 35 MRI_PER 0 Scanner is not online (might be switched off).
953 Time: 26.4.2018, 07:13:10, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (7044)
954E 26-Apr-2018 07:13:10 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
955 Time: 26.4.2018, 07:13:09, Process: CsaGenericMain (6904)
956E 26-Apr-2018 07:13:02 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection.
957 Time: 26.4.2018, 07:13:02, Process: CsaGenericMain (6632)
958E 26-Apr-2018 07:13:01 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
959 Time: 26.4.2018, 07:13:01, Process: CsaGenericMain (6632)
960E 26-Apr-2018 07:12:03 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
961E 26-Apr-2018 07:11:56 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
962E 26-Apr-2018 07:11:56 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
963E 26-Apr-2018 07:11:55 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
964E 26-Apr-2018 07:11:55 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
965E 26-Apr-2018 07:11:53 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
966E 26-Apr-2018 07:11:53 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
967E 26-Apr-2018 07:11:52 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
968E 26-Apr-2018 07:11:52 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
969E 26-Apr-2018 07:11:51 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
970E 26-Apr-2018 07:11:51 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
971E 26-Apr-2018 07:11:43 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
972E 26-Apr-2018 07:11:43 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
973E 26-Apr-2018 07:11:42 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
974E 26-Apr-2018 07:11:41 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\BRAdmin'.
975E 26-Apr-2018 07:11:34 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
976E 26-Apr-2018 07:11:32 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'NT AUTHORITY\SYSTEM'.
977E 26-Apr-2018 07:11:06 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Services\TmFilter\Parameters' with value 'Cue' by process C:\Program Files (x86)\Trend Micro\OfficeScan Client\NTRTScan.exe (Process Id: 2232, User: NT AUTHORITY\SYSTEM).
978E 26-Apr-2018 07:11:05 20 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify file 'c:\program files\mcafee\solidcore\sadmin.exe' by process C:\WINNT\system32\csrss.exe (Process Id: 468, User: NT AUTHORITY\SYSTEM).
979E 26-Apr-2018 07:11:05 20 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify file 'c:\program files\mcafee\solidcore\sadmin.exe' by process C:\WINNT\system32\csrss.exe (Process Id: 468, User: NT AUTHORITY\SYSTEM).
980E 26-Apr-2018 07:10:54 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'ctfmon.exe' by process C:\WINNT\SysWOW64\ctfmon.exe (Process Id: 1844, User: MRC31210\meduser).
981E 26-Apr-2018 07:10:54 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'internat.exe' by process C:\WINNT\system32\ctfmon.exe (Process Id: 1760, User: MRC31210\meduser).
982E 26-Apr-2018 07:10:54 49 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_USERS\S-1-5-21-4152650387-1435597237-583893964-1007\Software\Microsoft\Windows\CurrentVersion\Run' with value 'ctfmon.exe' by process C:\WINNT\system32\ctfmon.exe (Process Id: 1760, User: MRC31210\meduser).
983E 26-Apr-2018 07:10:46 50 McAfee Solidi... n/a McAfee Solidifier prevented an attempt to modify Registry key 'HKEY_LOCAL_MACHINE\SOFTWARE\Classes\Wow6432Node' by process C:\WINNT\system32\csrss.exe (Process Id: 468, User: NT AUTHORITY\SYSTEM).
984E 25-Apr-2018 23:00:19 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
985E 25-Apr-2018 23:00:18 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
986E 25-Apr-2018 23:00:17 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
987E 25-Apr-2018 23:00:17 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
988E 25-Apr-2018 22:59:28 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
989E 25-Apr-2018 22:59:16 32 MRI_COS 5 Connection to Image Reconstruction System has been lost.
990 Time: 25.4.2018, 22:57:20, Line: 876, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4840)
991E 25-Apr-2018 22:59:16 11 MRI_COS 4 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords.
992 Time: 25.4.2018, 22:57:11, Line: 838, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4840)
993E 25-Apr-2018 22:59:16 11 MRI_COS 2 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords.
994 Time: 25.4.2018, 07:14:08, Line: 793, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4840)
995E 25-Apr-2018 22:59:16 11 MRI_COS 1 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords.
996 Time: 25.4.2018, 07:14:07, Line: 838, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4840)
997E 25-Apr-2018 22:59:16 32 MRI_COS 0 Connection to Image Reconstruction System has been lost.
998 Time: 25.4.2018, 07:14:07, Line: 862, File: \n4\pkg\MrServers\MrVista\Cocos\CocosBE\CocosComp\Cocos.cpp, Process: CsaGenericMain (4840)
999E 25-Apr-2018 22:59:06 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details.
1000 Time: 25.4.2018, 22:59:06, Line: 156, File: \n4\pkg\MrApplications\MrExamUI\MrOnlineDisplay\MrDisplayerSvc\MrDisplayerDBSession.cpp, Process: CsaGenericMain (4316),
1001 Text: Failed to end session
1002 [LOG] An error occurred during accessing db. See logfile for details.
1003E 25-Apr-2018 22:59:06 96 McAfee Solidi... n/a McAfee Solidifier prevented package modification by 'c:\medcom\versant\vds605\lib\error.msi' by user: 'MRC31210\meduser'.
1004E 25-Apr-2018 22:59:04 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner.
1005 Time: 25.4.2018, 22:57:57, Process: CsaGenericMain (5328)
1006E 25-Apr-2018 22:58:36 35 MRI_PER 29 Scanner is not online (might be switched off).
1007 Time: 25.4.2018, 22:57:57, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1008E 25-Apr-2018 22:58:36 1 MRI_PER 28 Scanner is not online. Wait until booting has finished, or switch on Scanner.
1009 Time: 25.4.2018, 22:57:57, Process: CsaGenericMain (6612)
1010E 25-Apr-2018 22:58:36 15 MRI_PER 27 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
1011 Time: 25.4.2018, 22:57:57, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1012E 25-Apr-2018 22:58:36 2010 MRI_PXX 26 Patient table error: Voltage error (24V DC MDSD). Please call Siemens Service.
1013 Time: 25.4.2018, 22:57:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1014E 25-Apr-2018 22:58:36 2009 MRI_PXX 25 Patient table error: Voltage error (24V DC ORT). Please call Siemens Service.
1015 Time: 25.4.2018, 22:57:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1016E 25-Apr-2018 22:58:36 2008 MRI_PXX 24 Patient table error: Voltage error (24V DC SWT). Please call Siemens Service.
1017 Time: 25.4.2018, 22:57:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1018E 25-Apr-2018 22:58:36 9 MRI_STD 23 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
1019 Time: 25.4.2018, 22:57:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1020E 25-Apr-2018 22:58:36 207 MRI_SUP 22 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
1021 Time: 25.4.2018, 22:57:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1022E 25-Apr-2018 22:58:36 32 MRI_COH 19 Patient table error: Table position check failed Please retry the measurement. Please save System Log-Files immediately by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service.
1023 Time: 25.4.2018, 22:54:05, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1024E 25-Apr-2018 22:58:36 3000 MRI_PXX 13 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service
1025 Time: 25.4.2018, 22:49:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1026E 25-Apr-2018 22:58:36 440 MRI_SUP 12 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service.
1027 Time: 25.4.2018, 22:49:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1028E 25-Apr-2018 22:58:36 430 MRI_SUP 11 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service.
1029 Time: 25.4.2018, 22:49:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6612)
1030E 25-Apr-2018 22:58:26 35 MRI_PER 30 Scanner is not online (might be switched off).
1031 Time: 25.4.2018, 22:57:57, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1032E 25-Apr-2018 22:58:26 15 MRI_PER 29 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
1033 Time: 25.4.2018, 22:57:57, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1034E 25-Apr-2018 22:58:26 2010 MRI_PXX 28 Patient table error: Voltage error (24V DC MDSD). Please call Siemens Service.
1035 Time: 25.4.2018, 22:57:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1036E 25-Apr-2018 22:58:26 2009 MRI_PXX 27 Patient table error: Voltage error (24V DC ORT). Please call Siemens Service.
1037 Time: 25.4.2018, 22:57:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1038E 25-Apr-2018 22:58:26 2008 MRI_PXX 26 Patient table error: Voltage error (24V DC SWT). Please call Siemens Service.
1039 Time: 25.4.2018, 22:57:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1040E 25-Apr-2018 22:58:26 9 MRI_STD 25 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
1041 Time: 25.4.2018, 22:57:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1042E 25-Apr-2018 22:58:26 207 MRI_SUP 24 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
1043 Time: 25.4.2018, 22:57:44, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1044E 25-Apr-2018 22:58:26 32 MRI_COH 21 Patient table error: Table position check failed Please retry the measurement. Please save System Log-Files immediately by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service.
1045 Time: 25.4.2018, 22:54:05, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1046E 25-Apr-2018 22:58:26 3000 MRI_PXX 15 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service
1047 Time: 25.4.2018, 22:49:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1048E 25-Apr-2018 22:58:26 440 MRI_SUP 14 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service.
1049 Time: 25.4.2018, 22:49:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1050E 25-Apr-2018 22:58:26 430 MRI_SUP 13 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service.
1051 Time: 25.4.2018, 22:49:08, Line: 151, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1052E 25-Apr-2018 22:57:57 35 MRI_PER 0 Scanner is not online (might be switched off).
1053 Time: 25.4.2018, 22:57:57, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1054E 25-Apr-2018 22:57:57 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service.
1055 Time: 25.4.2018, 22:57:57, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1056E 25-Apr-2018 22:57:44 2010 MRI_PXX 0 Patient table error: Voltage error (24V DC MDSD). Please call Siemens Service.
1057 Time: 25.4.2018, 22:57:44, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1058E 25-Apr-2018 22:57:44 2009 MRI_PXX 0 Patient table error: Voltage error (24V DC ORT). Please call Siemens Service.
1059 Time: 25.4.2018, 22:57:44, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1060E 25-Apr-2018 22:57:44 2008 MRI_PXX 0 Patient table error: Voltage error (24V DC SWT). Please call Siemens Service.
1061 Time: 25.4.2018, 22:57:44, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1062E 25-Apr-2018 22:57:44 9 MRI_STD 0 Online Stimulation Monitor error: Gradient power amplifier was switched off, because gradient stimulation limits have been exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem persists: Please, contact service with the hint that details are available in service level.
1063 Time: 25.4.2018, 22:57:44, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1064E 25-Apr-2018 22:57:44 207 MRI_SUP 0 Scanner hardware error: Failure of the AMC power supply power. If the problem still persists, please call Siemens Service.
1065 Time: 25.4.2018, 22:57:44, Line: 170, File: \n4\pkg\MrCommon\MrNFramework\MrNlsMsg\MRNlsMessageId.cpp, Process: CsaGenericMain (6664)
1066E 25-Apr-2018 19:46:47 7 MRI_MIF 1 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service.
1067 Time: 25.4.2018, 19:46:47, Line: 1036, File: \n4\pkg\MrServers\MrMeasSrv\MeasCtrl\BE\MeasIFGC.cpp, Process: CsaGenericMain (6248),
1068 Text: SPRINT error
1069E 25-Apr-2018 17:54:57 8 MRI_RFS 0 SAR limit(s) exceeded.
1070 Time: 25.4.2018, 17:54:57, Process: CsaGenericMain (6248)
1071E 25-Apr-2018 17:54:57 21 MRI_MSR 1 Preparation of Measurement System failed.
1072 Time: 25.4.2018, 17:54:57, Line: 3558, File: \n4\pkg\MrServers\MrMeasSrv\MesSer\BE\MesSerCtrlImpl.cpp, Process: CsaGenericMain (6248),
1073 Text: CMD_PREP failed
1074E 25-Apr-2018 17:54:57 8 MRI_RFS 0 SAR limit(s) exceeded.
1075 Time: 25.4.2018, 17:54:57, Line: 2344, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6248),
1076 Text: ERROR: Prediction for SAR-Values: Limits exceeded
1077E 25-Apr-2018 17:54:57 8 MRI_RFS 0 SAR limit(s) exceeded.
1078 Time: 25.4.2018, 17:54:57, Line: 2344, File: \n4\pkg\MrServers\MrPatientSafetySrv\RFSWD\RFSWD_Tracer\RFSWDTracer.cpp, Process: CsaGenericMain (6520),
1079 Text: ERROR: Prediction for SAR-Values: Limits exceeded